Nuclear Medicine practice questions and answers 2026. Tap an option to test yourself — you'll see the correct answer and a plain-English explanation for every question. Free, no login.
Q11Which reconstruction algorithm is most commonly used for clinical SPECT and incorporates noise suppression via a filter?
✓ Correct answer: C. Filtered back-projection (FBP)
Filtered back-projection applies a ramp filter (removes star artifact) combined with a smoothing window (e.g., Butterworth or Hann) to suppress noise. While OSEM/MLEM are increasingly used, FBP with filtering remains the classic standard reconstruction in SPECT.
Topic: Instrumentation
Q12In OSEM (Ordered Subsets Expectation Maximization) reconstruction, one 'iteration' of OSEM with 8 subsets is equivalent to how many iterations of the standard MLEM algorithm?
✓ Correct answer: D. 8
Each OSEM iteration processes all subsets sequentially; with S subsets, one OSEM iteration approximates S MLEM iterations. Therefore 1 OSEM iteration with 8 subsets ≈ 8 MLEM iterations, making OSEM ~8× faster to converge.
Topic: Instrumentation
Q13Which method is the gold standard for attenuation correction in clinical SPECT/CT?
✓ Correct answer: B. CT-based attenuation correction using Hounsfield unit conversion
CT-based attenuation correction converts Hounsfield units to linear attenuation coefficients at the imaging energy, providing patient-specific, high-resolution maps. This has replaced external transmission sources in modern SPECT/CT systems.
Topic: Instrumentation
Q14PET imaging detects pairs of 511 keV photons. What physical process produces these photons?
✓ Correct answer: A. Positron-electron annihilation
When a positron emitted from a PET radionuclide travels a short distance and annihilates with an electron, the combined mass is converted into two 511 keV photons emitted 180° apart. PET cameras detect these coincident photons to localize the event.
Topic: Instrumentation
Q15LSO (lutetium oxyorthosilicate) is preferred over BGO in modern PET scanners primarily because of its:
✓ Correct answer: D. Much shorter scintillation decay time enabling time-of-flight capability
LSO has a decay time of ~40 ns vs. BGO's ~300 ns. The fast decay enables time-of-flight (TOF) PET, where the difference in photon arrival times localizes the annihilation event more precisely along the line of response, improving image signal-to-noise ratio.
Topic: Instrumentation
Q16Which scanner geometry parameter most increases PET system sensitivity?
✓ Correct answer: C. Increasing axial field of view (longer detector rings)
Sensitivity scales with the solid angle subtended by the detector, which increases with axial FOV (longer detector rings). Widening the energy window accepts more scatter; widening the time window increases randoms. Axial FOV extension is the dominant geometry-based sensitivity gain.
Topic: Instrumentation
Q17In PET, 'random coincidences' arise from two unrelated annihilation events detected within the coincidence time window. How are randoms typically corrected?
✓ Correct answer: B. Subtracted using a delayed coincidence channel
The delayed coincidence method introduces a time delay (e.g., 128 ns) in one detector channel so that any coincidences measured must be random (true coincidences cannot occur across this delay). These randoms-only counts are subtracted from the prompt channel.
Topic: Instrumentation
Q18For a PET scanner with a singles rate of S = 2 × 10⁶ cps per detector and a coincidence time window of 2τ = 10 ns, what is the approximate random coincidence rate R per detector pair?
✓ Correct answer: A. 0.04 cps
R = 2τ × S₁ × S₂ = 10 × 10⁻⁹ × (2 × 10⁶)² = 10 × 10⁻⁹ × 4 × 10¹² = 0.04 cps per detector pair. Random rates increase with the square of singles rate and are minimized by narrow time windows.
Topic: Instrumentation
Q19Daily quality control of a gamma camera should include which test?
✓ Correct answer: C. Peaking (energy calibration) and flood uniformity
NEMA and ACR standards specify daily peaking (verifying the photopeak window is correctly set) and flood-field uniformity (checking detector response homogeneity) as the minimum daily QC for gamma cameras.
Topic: Instrumentation
Q20A gamma camera flood uniformity image shows a 'ring' artifact on SPECT but not on planar images. The most likely cause is:
✓ Correct answer: D. Incorrect center-of-rotation (COR) offset
COR error causes each point source to be back-projected in a small circle rather than a point, producing ring or halo artifacts in reconstructed SPECT images. These rings are not visible in planar images because back-projection is not applied there.
Topic: Instrumentation
Use the free Nuclear Medicine Exam Prep sample, download the PDF, then unlock web-based timed mock exams for a full exam rehearsal.